OUTPUT RATINGS |
CH1/CH2 Independent | 0 ~ 30V / 0 ~ 6A ; 0 ~ 60V / 0 ~ 3A |
CH1/CH2 Series | 0 ~ 60V / 0 ~ 6A ; 0 ~ 120V / 0 ~ 3A |
CH1/CH2 Parallel | 0 ~ 30V / 0 ~ 12A ; 0 ~ 60V / 0 ~ 6A |
CH3 | 0.1 ~ 5V / 3A |
VOLTAGE REGULATION |
Line | ≦0.01% + 3mV |
Load | ≦0.01% + 5mV (rating current ≦6A) ≦0.01% + 8mV (rating current ≦12A) |
Ripple & Noise | ≦5mVrms (5Hz ~ 1MHz ) ≦ 50mVpp (20Hz ~ 20MHz) |
Recovery Time | ≦100μs(50% load change, minimum load 0.5A) |
CURRENT REGULATION |
Line | ≦0.2% + 3mA |
Load | ≦0.2% + 3mA |
Ripple & Noise | ≦3mArms |
TRACKING OPERATION |
Tracking Error | ≦0.5% + 10mV of master |
Series Regulation | ≦300mV |
Ripple & Noise | ≦10mVrms (5Hz ~ 1MHz) ≦100mVpp (20Hz ~ 20MHz) |
OUTPUT ON/OFF RESPONSE TIME |
Voltage Up (10% ~ 90%) | ≦100ms ( ≦ 95% rating load) |
Voltage Down (90% ~ 10%) | ≦100ms ( ≧ 10% rating load) |
OVP |
Accuracy | ± (0.5% of reading + 0.5V) |
METER |
Type | 3 1/2 digits 0.5" LED display |
Accuracy | ± (0.5% of reading + 2 digits) |
Resolution | 100mV/10mA |
INSULATION |
Chassis & Terminal | 100MΩ or above (DC 1000V) |
Chassis & AC code | 100MΩ or above (DC 1000V) |
TEMPERATURE COEFFICIENT |
Voltage | ≦ 100ppm/°C + 3mV |
Current | ≦ 150ppm/°C + 3mA |
REMOTE CONTROL |
| Output On/Off |
FAN NOISE |
| ≦50dB |
OPERATION ENVIRONMENT |
| Ambient temperature 0 ~ 40℃ Relative humidity ≦ 80% |
STORAGE ENVIRONMENT |
| Ambient temperature -10 ~ 70℃ Relative humidity ≦70% |
POWER SOURCE |
| AC 115V/230V ±15%, 50/60Hz |
ACCESSORIES |
| User manual x 1 Power code x 1 Test lead GTL-104 x 2, GTL-105 x 1 or European test lead GTL-201 x 1, GTL-203 x 1, GTL-204 x 2 |
DIMENSIONS & WEIGHT |
| 255 (W) x 145 (H) x 265 (D) mm Approx. 6kg |